Disassembled Ipas MVA Plus
Cleaning Ipas Instruments
Wash all surfaces thoroughly in warm water and detergent. Detergent is preferable to soap, which can
leave a residue.
CAUTION: Do not use any pointed or sharp objects to clean the valve parts or to remove the O-ring.
This could cause damage and prevent the device from maintaining vacuum.
Common Processing Methods for Ipas Instruments
After cleaning, the Ipas MVA Plus and Single-Valve aspirators, Ipas EasyGrip cannulae and adapters
(if used) must undergo high-level disinfection (HLD) or sterilization between patients to remove
contaminants. Devices are then safe to use for the next procedure. Aspirators and adapters do not need
to remain high-level disinfected or sterile for the next use. Cannulae must be high-level disinfected or
sterile at the time of use.
CAUTION: It is very important to follow these guidelines to ensure proper processing and to avoid
damage to the instruments.
valve
liner
plunger
Disassembled Ipas Single-Valve
cylinder
valve
plunger O-ring
collar stop
Aspirator