6. Calibration

CALIBRATION ADJUSTMENT

The furnace you received is precisely pre-calibrated at the factory. But in order to enable
even more flexibility, the CALIBRATION ADJUSTMENT feature has been introduces. It
enables you to adjust the muffles temperature to the level needed without performing
additional calibration (in case the discrepancy is small).
NOTE:
The new furnace, when it comes from the manufacturer, is properly
calibrated and checked.
In order to perform the calibration adjustment please perform the following:
Press Menu
Then press Options
Press arrow down until you reach the + of the " Temp calibration"
parameter high light
Press "ENTER" key to modified "+" to "-".
Press arrow down to enter to the data field .In the DATA field enter
the numeric value
Press "menu": or "Ent " to validate your change
The change will affect the firing parameters in all programs.
The change will be stored in the memory and all the firing temperature will be higher or
lower accordingly.
In order to return to the basic, factory made calibration perform the
following:
Enter to the calibration adjustment mode again and cancel the stored value
Calibration will be restore
FOCUS 4010/HT Dental Porcelain Furnace
Page 21
Manual version 20.08.2015