OPERATING ISSUE
Unit will not turn on.
Cannot zero the scale, or will not
zero when turned on.
Unable to calibrate.
Cannot display weight in
desired weighing unit.
Cannot change menu settings.
Error 8.1
Error 8.2
Error 8.3
Error 8.4
Error 8.6
Error 9.5
CAL E
NO.SW
REF. WT Err
If the troubleshooting section does not resolve your problem, contact ULINE Customer Service at 1-800-295-5510.
π
CHICAGO • ATLANTA • DALLAS • LOS ANGELES • MINNEAPOLIS • NYC/PHILA • SEATTLE • MEXICO • CANADA
PAGE 10 OF 30
TROUBLESHOOTING
CAUSES
Power cord not plugged in or properly connected.
Power outlet not supplying electricity.
Other failure.
Load on scale exceeds allowable limits.
Load on scale is not stable.
Load cell damage.
Lock Calibration Menu set to On.
LFT menu set to On.
Incorrect value for calibration mass.
Unit not set to On.
Menu has been locked.
Weight reading exceeds Power On Zero limit.
Weight reading below Power On Zero limit.
Weight reading exceeds Overload limit.
Weight reading below Underload limit.
Weight exceeds six digits. Display overflow.
Calibration data not present.
Calibration value outside allowable limits.
Attempting to exit the menu with the LFT setting ON
and the security switch OFF.
Reference Weight too small. The weight on the
platform is too small to define a valid reference
weight.
1-800-295-5510
Check power cord connections. Make sure power
cord is properly plugged into the power outlet.
Check power source.
Service required.
Remove load on scale.
Wait for load to become stable.
Service required.
Set Lock Calibration Menu to Off.
Set LFT Menu to Off.
Use correct calibration mass.
Enable units in the Units Menu.
Set selected menu to Off in the Lock Menu.
Lockout Switch on the circuit board may need to
be set to the Off position.
Remove load from scale. Recalibrate scale.
Add load to scale. Recalibrate scale.
Reduce load on scale.
Add load to scale. Recalibrate scale.
Reduce load on scale.
Calibrate scale.
Use correct calibration weight.
Refer to Settings in Legal for Trade section, page 7.
Set the security switch to the ON position.
Use a greater weight for sample.