7. CLEANING

8. WHAT TO DO IF ...?

Malfunction
... no weight is displayed
during weighing?
0. 0 0
...
does not appear
before weighing?
... a segment lights up
continuously or not at
all?
...
is displayed?
bAtt
...
is displayed?
St0p
...
is displayed?
60 •
NOTE:
For further operation of the wireless printer see
its user manual.
8. If you do not wish to make any more settings, press
the Start key.
The settings will be saved.
The device switches off.
To clean the covering and housing use a household
detergent or standard disinfectant as required. Take
note of the instructions provided by the manufacturer.
Do not use scouring or caustic cleaning products,
spirit, petrol or similar substances for cleaning under
any circumstances. Such products may damage the
high-quality finishes.
Cause/Remedy
The scales are not connected to the power
supply
- Check the scales are switched on
- Check batteries are inserted
A load was placed on the scales before being
switched on.
- Remove load
- Switch scales off and back on again
There is a fault at that point.
- Notify maintenance service.
Battery voltage is falling off.
- Insert new batteries as soon as possible
Batteries are flat.
- Insert new batteries
The maximum load has been exceeded.
- Reduce load