Restoring Factory Settings; Hygiene Treatment; Cleaning - Seca 655 Manual De Instrucciones

Ocultar thumbs Ver también para 655:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Restoring factory settings

8. HYGIENE TREATMENT

8.1 Cleaning

1. Press the
key.
The Settings menu is displayed.
2. Press the
or
displayed.
3. Press the Reset item.
4. Press the Factory settings key.
The device will be reset to factory settings.
The following network settings are not reset:
• Server address
• Server port
• WiFi
• Access point
5. Confirm your selection.
6. Confirm the prompt on the display with the
You exit the menu automatically.
The device will be reset to factory settings.
WARNING!
Electric shock
The device is not de-energized when the display goes out. Use of
fluids on the device may cause electric shock.
► Disconnect the power supply connector before performing any
hygiene treatment.
► Ensure that no fluids penetrate the device.
NOTICE!
Damage to device
Inappropriate detergents and disinfectants may damage the sensitive
surfaces of the device.
► Use alcohol-based disinfectant (e.g. 70 % ethanol) exclusively.
► Do not use aggressive or abrasive cleaning agents.
► Ensure that no moisture or dust get into the sensors during
cleaning.
► Use a soft cloth dampened with mild soapsuds to clean the surfaces of
the device.
key until the Voice guidance menu item is
key.
Hygiene treatment • 123