Troubleshooting - Uline H-4597 Instrucciones De Funcionamiento

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

7.
Enter the Setup menu.
8. Set the desired Filter level.
9. Set the desired Power on Unit setting.
10. Set LFT to ON.
11. Set the Security Switch to ON. (See Figure 2)
12. Exit the Setup menu.
VERIFICATION
The local weights and measures official or authorized
service agent must perform the verification procedure.
SEALING
The local weights and measures official or authorized
service agent must apply a security seal to prevent
tampering with the settings. Refer to Figure 3 for sealing
methods.
OPERATING ISSUE
Cannot turn on.
Poor accuracy.
Cannot calibrate.
Cannot access mode.
Cannot access unit.
Err 8.1
Err 8.2
OVER
UNDER
Err 9
REF Err
LOWrEF
If the problem persists, or if the troubleshooting section does not resolve or describe your problem, contact
Uline Customer Service at 1-800-295-5510.
PAGE 4 OF 17
OPERATION CONTINUED

TROUBLESHOOTING

CAUSES
No power to scale.
Improper calibration.
Unstable environment.
Unstable environment.
Incorrect calibration weight.
Mode not enabled.
Unit not enabled.
Pan has load during power on.
Transport lock is in locked position.
Pan was removed prior to power on.
Transport lock is in locked position.
Weight on pan exceeds capacity.
Pan was removed during weighing.
Factory calibration data corrupted.
Reference weight is too small.
Reference weight is too low for
accurate percent weighing.
Figure 3
WIRE SEAL:
Remove the spacers under the
cross-drilled screws. Reinstall
the screws and feed the wire
through the cross-drilled holes.
RECOMMENDATIONS
Verify connections and voltage.
Perform calibration.
Move scale to suitable location.
Move scale to suitable location.
Use correct calibration weight.
Enter menu and enable mode.
Enter menu and enable unit.
Remove weight from pan and re-zero.
Unlock transport lock.
Install pan and re-zero.
Unlock transport lock.
Remove weight from the pan.
Reinstall pan.
Contact Ohaus or an authorized
service agent. Perform service
calibration.
Use larger sample.
Continue to weigh with less accurate
results.
PAPER SEAL:
Place the paper seals over the
screws and sealing spacers.
0721 IH-4597