Chlorine T / M100
Implementation of the provision total Chlorine with tablet
Select the method on the device
In addition, choose the test: total
For this method, no ZERO measurements are to be carried out with the following de-
vices: XD 7000, XD 7500
Fill 24 mm vial with 10 ml
sample.
Zero
Press the ZERO button.
For devices that require no ZERO measurement , start here.
Add DPD No. 1 tablet .
11
10 ml
Close vial(s).
Remove the vial from the
sample chamber.
Add DPD No. 3 tablet .
Sample
Place sample vial in the
sample chamber. • Pay at-
tention to the positioning.
Empty vial except for a few
drops.
24 mm
Crush tablet(s) by rotating
slightly.
EN   Method Reference Book   1.0