Chlorine T / M100
Implementation of the provision Chlorine differentiated with
tablet
Select the method on the device
In addition, choose the test: differentiated
For this method, no ZERO measurements are to be carried out with the following de-
vices: XD 7000, XD 7500
Fill 24 mm vial with 10 ml
sample.
Zero
Press the ZERO button.
For devices that require no ZERO measurement , start here.
Add DPD No. 1 tablet .
13
10 ml
Close vial(s).
Remove the vial from the
sample chamber.
Crush tablet(s) by rotating
slightly.
Place sample vial in the
sample chamber. • Pay at-
tention to the positioning.
Empty vial except for a few
drops.
24 mm
Fill up vial with sample to
the 10 ml mark.
EN Method Reference Book 1.0
Sample