Ac Current Measurement / A - Testboy TB 312 Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 75

AC current measurement / A~

Set the appropriate range using the selector switch. Connect the black test lead to the
"COM" jack and the red test lead to the mA or 10 A jack. Connect test leads to the test
specimen. Read off measurement result from the display.
You must use the "10 A" jack when measuring currents above
500 mA!
Alternating current
Measuring
range
500 µA
5000 µA
50 mA
500 mA
5 A
10 A*
Accuracy specified only valid for 10 % - 100 % of the measuring range.
- 500 mA range is protected by a F 500 mA / 1000 V fuse.
- 10 A range is protected by a F 10 A / 1000 V fuse.
- in the 10 A range observe the maximum duty cycles!
- Frequency range: 40 – 20000 Hz.
* Following measurement, which should take a
max. 10 seconds, allow the instrument to cool down
for 15 minutes to protect against overheating.
Press the SELECT button to switch between DC, AC and AC+DC measurement
methods.
®
Testboy
TB 312
Resolution
40 Hz – 1 kHz
± 0.75 % of
0.01 µA
meas. val.
+ 20 digits
± 0.75 % of
0.1 µA
meas. val.
+ 10 digits
± 0.75 % of
1 µA
meas. val.
+ 20 digits
± 0.75 % of
10 µA
meas. val.
+ 10 digits
± 0.75 % of
0.1 mA
meas. val.
+ 20 digits
± 1 % of
1 mA
meas. val.
+ 10 digits
Accuracy
1 kHz – 10 kHz
10 kHz – 20 kHz
± 1 % of
± 1.2 % of
meas. val.
meas. val.
+ 20 digits
+ 5 digits
± 1 % of
± 1.2 % of
meas. val.
meas. val.
+ 10 digits
+ 5 digits
± 1 % of
± 1.2 % of
meas. val.
meas. val.
+ 20 digits
+ 5 digits
± 1 % of
± 1.2 % of
meas. val.
meas. val.
+ 10 digits
+ 5 digits
± 1.5 % of
± 1.2 % of
meas. val.
meas. val.
+ 20 digits
+ 5 digits
± 1.5 % of
not specified
meas. val.
+ 10 digits
Operation
41

Publicidad

Tabla de contenido
loading

Tabla de contenido