Installation
The Transducer Spray Cleaning Nozzle attaches quickly and easily to your application.
1.
Attach 3/4" conduit pipe to transducer using a 3/4" NPT to 1" NPT coupling. (See
maximum height below.)
2.
Run cable through conduit, and make sure that the conduit pipe extends above the
Nozzle bracket.
3.
Attach Nozzle to conduit using the supplied SS U-bolts. Do not tighten.
4.
Move Nozzle down until the bracket touches the transducer. (As in detail view
below.)
5.
Tighten U-bolts.
6.
Attach water source to 3/4" Nozzle connection.
7.
Regulate frequency and pressure as required.
move nozzle
down until it
touches the
transducer
Page 2
Transducer Spray Cleaning Nozzle – INSTRUCTION MANUAL
nozzle bracket
70 mm (2.75") max
7ML19985HH61