Using the SFB Serial Interface

The SFB serial connection on the console rear panel enables FireWire connection
to the Stryker Endoscopy Software Management Site (SMS). Connecting to this
site enables remote diagnostics and software updates.
Note
This system feature is not necessary for regular camera system
operation.
Note
This system feature requires an additional device (that is, a
computer) to connect to SMS.

Using the DVI Fiber Outputs

The 1188HD Camera contains four laser diodes to transmit a DVI output over
fiberoptic cables.
1.
Connect four individual fibers (terminated in Lucent connectors) to the
red (R), green (G), blue (B) and clock (C) laser diodes on the console
rear panel.
2.
Connect the four fibers to a compatible fiberoptic DVI receiver.
• The four fibers should be connected to the camera console in the
labeled order: RGBC
• The fibers should be connected to the monitor in one of two
configurations: CBGR (reverse order) or BGRC (R/B switched).
3.
Ensure the fiberoptic output is enabled (on) via the Options Submenu in
the Configuration Menu. (See the section "Using the Configuration
Menu".)
Note
The 1188HD Camera is a Class 1 laser product per IEC 60825-1
and 21CFR.
Warning
Using controls or adjustments or performing procedures
differently than specified in this manual may result in
hazardous radiation exposure.