Condition
OCC OUT
• Calibration not yet completed
(repeated)
LOCK
• Only the current feeding
ER01 - ER99
• The self test detected an electronic error
ER40
• Keypad pressed or stuck during pump
No plug symbol visible,
• The wall outlet doesn't work.
while the pump is con-
nected to the mains.
• The charger is damaged.

SET UP MODE

The set up mode is used
This set up mode may only be entered by healthcare professionals and authorised personnel trained to use this application of
the pump.
The patient or his/her relatives without permission of the physician, dietician, nurse or other licensed practitioner, may not
change the settings of the set up mode.
+
ON/OFF
&
2 sec
2 sec
Hold the '+' key down and press "ON/OFF" for 2 seconds. The pump enters the set up mode.
First the alarm level can be set:
"BEEP HIGH" (+) or "BEEP LOW" (-)
Use the "+" or "-" key to change the setting from "high" to "low" or from "low" to "high".
ZMNL900911_AW.indd 82
ZMNL900911_AW.indd 82
Cause
on the current feeding set
schedule is allowed for this patient.
Another feeding program is not allowed
by your healthcare professional.
start up.
• to set the alarm level,
• to lock the keypad,
• to switch the audible alarm off when a dose is administered,
• to switch the light of the LCD permanently on when connected to the mains,
• to enable connection with a Nurse call or Patient Data
Monitoring System (PDMS).
TO SET THE ALARM LEVEL
+
Beep high
d
-
Beep low
Correction
• Stop the alarm by pressing the "START/STOP" key.
• Start the pump by pressing the "START/STOP" key and let it run
only briefly.
• Stop the pump by pressing the "START/STOP" key insuring that
there has been no occlusion out alarm.
• Remove the feeding set from the pump and reinsert the feeding set
in the pump.
• Restart the pump by pressing the "START/STOP" key.
• The programming feature is blocked in the set up mode of the
pump. Ask your healthcare professional to modify this setting.
• Turn the pump "off", make sure the pump door is closed and
switch the pump back "on". If the error persists, contact your
healthcare provider or refer the pump to service.
• Restart pump by ensuring keypad is not pressed.
• Connect the pump to another wall outlet.
• Contact your healthcare professional or
Nutricia subsidiary to replace the charger.