Reset patient data is designed to erase all collected patient data (non-personal)
The following data will reset: Therapy parameters, Progress bar, Patient history data, Personalised therapies, Define therapy list, Sound settings and Display settings.
5.1.8.
On/Off button
The ON/OFF button switches the pump ON and OFF.
Note: to avoid unintended button presses, the ON/OFF button needs to be pressed for 2 seconds for both switching the pump ON or OFF.
5.1.9. Soft buttons
The soft buttons relate to the function on the screen which has an action indicated by the "<, >, ,,,,+,-" symbol.
Note: the SOFT buttons are only active when accompanied with the above-mentioned symbols, if no symbol appears, the button is inactive.
5.2.
PUMP STATUS INDICATOR
The pump status indicator provides a visual indication of the actual pump status.
•
A green light indicates that the pump is in delivery mode and working normally.
•
An orange light indicates that the pump is in pause mode and waits for user input.
•
A red light indicates an alarm state; immediate action from the user is required.
6.
PUMP SETUP
6.1.
ATTACHING TO THE POLE CLAMP (OPTIONAL)
1.
Connect the Flocare Infinity III pole clamp (N) to a pole stand by rotating the black star shaped bolt clockwise.
2.
Connect the Flocare Infinity III to the pole clamp interface (M) by rotating the grey wheel shaped disk clockwise until the pole clamp is fully connected to the
pump without any clearance.
6.2.
CONNECT TO THE AC POWER PLUG [e-IFU]
Connect the AC power adapter by gently pressing the charger plug (E) into the pump (J).
The charger plug is designed to fit in only one way. When the pump is connected to the mains, the blue AC charging light (K) should be on and the plug icon to be
appeared next to the battery icon.
Note: The Flocare Infinity III battery needs to be fully charged for a period of 6 hours prior to or during first use.
6.3.
SWITCH THE PUMP ON [e-IFU]
Turn the pump ON by pressing and holding the ON/OFF button for approximately 2 seconds until the start-up sound can be heard. The pump will perform a self-test
and inform the user when the device is ready to use, the home screen will be shown after the self-test.
Caution: Carefully check for any sections or pixels that do not appear during the display test. In case of missing sections or pixels, the pump should not be used and
send to an authorised service centre for inspection.
6.4.
SET THE LANGUAGE
At first pump start up or after a factory reset, the pump will guide you automatically through the language selection. Set the language by pressing the SETTINGS
button, navigate to LANGUAGE and select your preferred language.
6.5.
SET THE TIME AND DATE [e-IFU]
At first pump start up or after a factory reset, the pump will guide you automatically through the date and time setting. Set the date and time by pressing the
SETTINGS
button, navigate to TIME AND DATE and set these to your local time zone.
6.6.
SWITCH THE PUMP OFF
Turn the pump OFF by pressing and holding the ON/OFF button for approximately 2 seconds until you hear the turn-off sound.
7.
OPERATING INSTRUCTIONS [e-IFU]
7.1.
LOAD THE FLOCARE INFINITY GIVING SET
Follow the instructions for use enclosed with the Flocare Infinity giving set, for setting up the enteral system.
1.
Open the pump door
2.
Position the purple looped section of the cassette (R) around the pump rotor (D)
3.
Gently stretch (do not over stretch) and seat the cassette into the pump The cassette is designed to fit into the pump one way
4.
Close the pump door
5.
Pump is now ready to be programmed
7.2.
PAUSE / RUNNING MODE [e-IFU]
The Flocare Infinity III is either in pause mode waiting for user instructions or it is in running mode delivering a therapy to a patient. The pause or running modes are
controlled by the START/PAUSE [
7.3.
FILL THE FLOCARE INFINITY GIVING SET [e-IFU]
The FILL SET function is used to prime a (new) Flocare Infinity giving set with nutrition or water and needs to be commenced before starting a feeding therapy.
Ensure the Flocare Infinity giving set is disconnected from the patient feeding tube. Prime the set by press and hold the FILL SET
the FILL SET function and release the FILL SET
by pressing STOP in case the giving set is over filled.
ZMNL900901_DE.indb 8
ZMNL900901_DE.indb 8
] button.
button as soon as the pump generates a beep and automatically starts. The FILL SET can be manually stopped
‒ 8 ‒
button for 2 seconds to activate
15.07.21 10:12
15.07.21 10:12
7.4.
'Define
with a
7.4.1.
Reset
The fo
7.4.2.
The Fl
For sa
The TH
warnin
7.5.
Cautio
Therap
Volu
Rate
Inter
IM
Th
the
7.5.1.
•
•
•
•
•
•
•
•
7.5.2.
•
•
•
•
7.6.
When
You ei
7.7.
7.7.1.
The Fl
The da
naviga
7.7.2.
The P
When
7.8.
The Fl
maxim
Note:
rechar