To Lock The Keypad - Nutricia Flocare Infinity Instrucciones De Uso

Ocultar thumbs Ver también para Infinity:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 31
UK-IE
+
Lock
FILL SET
d
-
Unlk
Hold the '+' key down and press "ON/OFF" for 2 seconds. The pump enters the set up mode. Press the "FILL SET" key to
switch to the next setting:
"UNLK" the keypad (-) or "LOCK" the keypad (+). In the "LOCK" mode the feeding program of the pump can no longer be
changed.
Use the "+" or "-" key to change the setting from "lock" to "unlk" or vice versa.
+
Beep when done
FILL SET
d
-
Mute when done
Hold the '+' key down and press "ON/OFF" for 2 seconds. The pump enters the set up mode. Press the "FILL SET" key to
switch to the next setting:
"BEEP WHEN DONE" (+): the pump will give an alarm when the dose is administered or "MUTE WHEN DONE" (-): the pump will
not give an alarm after administering the dose. Use the "+" or "-" key to change the setting from "beep when done" to "mute
when done" or vice versa.
+
Lite on
FILL SET
d
-
Lite off
Hold the '+' key down and press "ON/OFF" for 2 seconds. The pump enters the set up mode. Press the "FILL SET" key to
switch to the next setting:
"LITE ON" (+) the light of the LCD remains on when the pump is connected to the mains. "LITE OFF" (-) the light of the LCD
switches off after a few seconds even if connected to the mains. Use the "+" or "-" key to change the setting from "lite on" to "lite
off" or vice versa.
+
OUTP PDMS
FILL SET
d
-
OUTP NRSE
Hold the '+' key down and press "ON/OFF" for 2 seconds. The pump enters the set up mode. Press the "FILL SET" key to
switch to the next setting:
"OUTP PDMS" (+) to connect with a Patient Data Monitoring System. "OUTP NRSE" to connect with a NURSE CALL system. Use
the "+" or "-" key to change to setting from "OUTP PDMS", to "OUTP NRSE" or "OUTP OFF" and vice versa.
Press the "ON/OFF" key and hold it down for 2 seconds to exit the SET UP mode. The settings are automatically saved.
ZMNL900911_AW.indd 83
ZMNL900911_AW.indd 83

TO LOCK THE KEYPAD

TO MUTE WHEN DOSE DONE
TO SWITCH LIGHT ON
ON/OFF
d
2 sec
TO CONNECT WITH NURSE CALL OR PDMS
ON/OFF
d
2 sec
-83-