Technical data
en
Measurement accuracy
Standard deviation of the repeatability with ethyl alcohol standard
Up to 0.50 mg/L
≥0.50 mg/L
up to 0.100 % BAC
≥ 0.100 % BAC
Sensitivity drift
Typically 0.4 % of the measured value per month
Wait time for measurement readiness following a previous
measurement with:
0.25 mg/L
0.50 mg/L
1.00 mg/L
>1.50 mg/L
Calibration
every 12 months
Dimensions
1.96/2.36 " x 5.55" x 1.22" (W x H x D)
Weight
approx. 144 g/ 5.1 oz
Power supply
1 x 3 V CR123A
16
20 s
40 s
80 s
120 s
0.008 mg/L
1.7% of the measured value
0.0017 % BAC or
1.7% of the measured value
whichever value is higher
0.050 % BAC
0.100 % BAC
0.200 % BAC
>0.300 % BAC 120 s