Service information
Maintenance of a medical device is wholly the responsibility of the owner of that device. Failure to maintain a
device in accordance with the instructions may invalidate the device's warranty. Furthermore, failure to main-
tain a device may compromise the clinical condition or safety of users and/or their caregivers.
EN
1st Day
Read the User Guide carefully before use and
store it in a safe place for future reference
Store the enclosed tools on/at the product
Before every use
Use a cloth with warm water and a mild deter-
gent without chlorine to clean the product
Check that all castors are moving freely
and that all wheel locks/brakes are easily
activated
Make sure all cables are securely plugged in
before use in the shower
Inspect all cables for cracks or signs for worn-
out parts. Never use a product which appears
to be substandard or faulty. Contact your local
dealer for service
Disinfect the upholstery. please check the
disinfection section in this User Guide for
more information
Inspect the detachable parts. If required,
disinfect the parts with a 70% disinfectant
solution. It is recommended to wipe of any
residue and dirt using a cloth wrung in warm
water and a mild detergent/soap without
chlorine. Let the detachable parts dry before
disinfection
Daily
Use a dry cloth to clean the product
Visual inspection. Repair or replace any dam-
aged or worn-out parts
Ensure that all fasteners are in place and fit-
ted correctly
Check all fittings and buckles for signs of worn
out parts
Be aware of any MAX mark indications before
adjustment
Weekly
Check that all castors are moving freely and
that all wheel locks/brakes are easily activated
Wipe off any residue and dirt from the product
using a cloth with warm water and a mild deter-
gent/soap and dry before use
Monthly
All nuts and bolt on this product must be
checked and tightened to avoid unnecessary
failures
Oil the swing-away parts. We recommend you
use a professional lubricant system
ENGLISH
Annually
Inspect the frame for cracks or signs for worn-
out parts and carry out annual service. Never
use a product which appears to be substan-
dard or faulty
Inspect the piston rods are greased, extract the
piston rods to check grease
Warewashing machine
This product can be washed at:
85°C using a mild detergent in 3 minutes in a wash-
ing machine designed for medical devices. Use the
machine's drying function to dry the product. Clean
the accessories seperately.
60°C using a mild detergent in 15-20 minutes in a
washing machine designed for medical devices. Use
the machine's drying function to dry the product.
Clean the accessories seperately.
Before washing:
Avoid degreasing of the piston rods, the ac-
1.
tuators should be retracted to minimum stroke
and without load before washing.
Remove all accessories, unplug all cables and
2.
make sure to dismount the battery, control-
box and handset. Leave the actuators with the
cables mounted on the product (see page 24)
After washing:
Ensure correct placement of cables inserted
3.
into the control box when reinstalling the pow-
er unit system (see page 25)
Check that the handset functionalities works
4.
correct
Disinfection
The upholstery can be disinfected with a 70% disin-
fectant solution. It is recommended to wipe off any
residue and dirt from the upholstery, using a cloth
with warm water and a mild detergent/soap without
chlorine and let it dry before disinfection.
Service interval
A detailed inspection of the product must be made
every 12 months (every 6 months for heavy users)
and every time the product is reissued for use. The
inspection should be carried out by a technically
competent person who has been trained in the use
of the product.
The lifetime of this product in normal use is 5 years
if all maintenance and servicing is carried out in ac-
cordance with the manufacturer's instructions and
demonstrably recorded.
34