Malfunction
... if after switch-on,
measured results are
transmitted for the first
time and two acoustic
signals are heard?
... only the item "SYS" is
visible in the rf menu?
... only the items "SYS"
and "lrn" are visible in
the rf menu?
... the items "APrt" and
"Time" are not visible in
the rf menu?
Er:X:11
...
is displayed?
Er:X:12
...
is displayed?
Er:X:16
...
is displayed?
Cause/Remedy
• The device was unable to send measured
results to the radio receiver (seca wireless prin-
ter or PC with seca USB wireless module).
– Check that the scales are integrated in
the wireless network.
– Check that the receiver is switched on.
• Nearby RF equipment (e.g. mobile phones) are
interfering with reception.
- Make sure that RF equipment is kept at least
1 metre away from transmitters and receivers
in the seca wireless network.
NOTE:
If such interference is not eliminated, no
further acoustic warning will be given with
subsequent attempts at transmission.
• The wireless module is disabled.
- Activate wireless module (see "Activate
wireless module (SYS)" on page 52).
• The wireless module is activated but no wireless
group set up.
- Set up wireless group (see "Set up wireless
group (Lrn)" on page 52).
• No wireless printer is logged into the wireless
group.
- Use menu item "lrn" to log the wireless printer
into the wireless group (see "Set up wireless
group (Lrn)" on page 52).
The scales are too high, or too great a load is
applied at one point.
- Reduce load on scales or distribute weight
more evenly
- Restart scales
The scales have been switched on with too great
a load.
- Reduce load on scales
- Restart scales
Oscillation of the scales has occurred, preventing
determination of the zero point.
- Do not touch the trolley or table supporting
the scales
- Restart scales