5-37.Replacing the Fuses
Referring to Figure , examine or replace the Meter's fuses as follows:
1.Turn the Meter off and remove the test leads from the terminals.
2.Remove the battery door assembly by using a standardblade screwdriver to turn the battery
door screw one-half turn counterclockwise.
3.Remove the fuse by gently prying one end loose, then sliding the fuse out of its bracket.
4.Install only specified replacement fuses.
5.Reinstall the battery door assembly and secure it by turning the screw one-half turn clockwise.
Stand
battery cover
7.4V Li-ion battery
Stand
Seal ring of
battery cover
Fuse wire
5-38.Li-ion Battery Charge
1.Set the function switch to the OFF/CHG position.
2.Insert the socket into the Meter Input port. And
the Adapter connected to the switch socket.
Then Inser t the Adapter into Power socket.
3.Display charge symbol in TFT color LCD display.
24
battery cover
7.4V Li-ion
battery
Seal ring of
battery cover
Back shell
Seal ring of
back Shell
Fuse wire
Fuse wire
Adapter
Socket
13/08/2017 Version No. 001

6.General Specification

Enclosure
Double molded, waterproof
Shock (Drop Test)
6.5 feet (2 meters)
Diode Test
Test current of 0.9mA maximum, open circuit
voltage 3.2V DC typical
Continuity Check
Audible signal will sound if the resistance is less
than 25 (approx.), test current <0.35mA
PEAK
Captures peaks >1ms
Temperature Sensor
Requires type K thermocouple
Input Impedance
>10M VDC & >9M VAC
AC Response
True RMS
AC True RMS
The term stands for "Root-Mean-Square" which
represents the method of calculation of the voltage
or current value. Average responding multimeters
are calibrated to read correctly only on sine waves
and they will read inaccurately on non-sine wave
or distorted signals. True rms meters read
accurately on either type of signal.
ACV Bandwidth
50Hz to 100000Hz
< 3 at full scale up to 500V, decreasing linearly
Crest Factor
<
to
1.5 at 1000V
Display
50,000 count backlit liquid crystal with bargraph
Overrange indication
"OL" is displayed
Auto Power Off
5-30minutes (approximately) with disable feature
Polarity
Automatic (no indication for positive); Minus (-)
sign for negative
Measurement Rate
20 times per second
Low Battery Indication
" " is displayed if battery voltage drops below
operating voltage
13/08/2017 Version No. 001
MeterScope/English
25