Adjusting the cursor balance
The weight adjustment selector enables two different setting:
UP :
balance setting for hea-
vier probes (depth, 90°,
extended measuring
DOWN :
standard probe weight
setting
The weight adjustment selector enables with a screwdriver 00 or a pen.
Error messages
Err0
Sensor measuring error. Press SET, and control the measuring reference
Err1
Parity Error (RS232 | USB connexion)
Err2
Unkown command (RS232 | USB connexion)
Err6
Error during the probe measuring detection sensor procedure. In case of
a problem during Hall initialization, please proceed with following controls:
•
Is the probe free to move ?
•
Is the measuring force correctly set (see chapter « Selector adjustement ») ?
•
If REF1 and CMDE are displayed, UP/Down button may have been
activated during the initialization procedure. Please check that the UP /
DOWN button is free.
Possible action on the Keypad or FCT key during initialization.
Err7
Please check that keypad and FCT are free, then restart the Hi_Cal.
UP/Down may have been activated during the initialization procedure.
REF1
CMDE
Please check that the UP/DOWN button is free, and restart the Hi_Cal.