Adjusting the cursor balance
The weight adjustment selector enables two different setting:
101
UP :
balance setting for hea-
vier probes (depth, 90°,
extended measuring
DOWN :
standard probe weight
setting
31
The weight adjustment selector enables with a screwdriver 00 or a pen.
Matière:
Article:
Masse / Volume:
Error messages
Err0
A3
1/1
Err0.540
Err0.550
Err1
Err2
Err6
PrObE
22
- - -
M302-150-101
44513824299.110 g
Cotes non tolérancées selon SN EN 22768-1 / Classe f (fine), tolérances indiquées en 1/1000 mm
Ensemble général 150
Measurement error of the main sensor. Acknowledge via SET,
new reference tap.
CH-1400 Yverdon-les-Bains (Suisse)
Measurement error of the UP/DOWN button sensor. Release
and acknowledge via SET
Measurement error of the relative sensor. Release the mea-
suring key and acknowledge with SET.
Parity error in an ASCII RS232 | USB command
RS232 control | USB ASCII
Error when initializing the centering of the measuring range.
Check the following points and restart the instrument:
- Does the measuring key move freely?
- Is the selector switch for key weight compensation correctly set?
? (see chapter «Adjusting the selector switch»)»
Measuring key blocked. Release measuring key (when initiali-
zing the instrument)
Chanfreiner:
Trait. thermiques:
-
Trait. de surfaces:
-
Echelle
08-Jun-20
Dess.:
2:5
Modèle:302_150_101{.prt/asm}
Liste de pces selon nomenclature:
-
PM 302.150.101 -
DNi