Protected against solid foreign objects of 12.5mm
IP20
and greater. No protection against water.
Use by date
Lot/batch code
Sterilized using irradiation
For US market only: Prescription use only. U.S. Federal
RX-
Law restricts this device to sale by or on the order of a
Only
licensed practitioner
REF
Catalog Number
2
Do not reuse
7) Before You Start
The Osstell Beacon is delivered from factory in a "transport" mode,
where the built-in motion sensor is deactivated.
To de-activate the transport mode and start charging, connect the
small USB connector of the Osstell USB cable to wide end of the
instrument. Connect the large USB connector to a standard USB
type A outlet of a PC, laptop or charger.
The Osstell Beacon will start-up and enter charging mode. Charge
the Osstell Beacon for at least 3 hours or until the Osstell Beacon
indicates full charge. Remove the Osstell USB cable.
Note! It is not possible to perform an ISQ measurement during
charging.
8) Operation of the Osstell Beacon
Instrument
The Osstell Beacon is activated by a built-in motion sensor. As soon
as the motion sensor detects movement, the Osstell Beacon will
start-up, flash red-yellow-green and shortly show battery status
in the lower display and then be ready for measurement in the BL
(Buccal – Lingual) direction, which is indicated in the upper display.
The Osstell Beacon will start to measure, indicated by an audible
sound, when near an Osstell Smartpeg or Testpeg due to a built-in
magnetic detector that senses the magnet on the top of the Osstell
Smartpeg/TestPeg. Measured data will be shown in the upper display
combined with a colored light indication below the instrument tip.
The red, yellow and green color represents low, medium or high
implant stability. After 60 seconds of no motion, the Osstell Beacon
will turn off automatically.
9) TestPeg
The Osstell TestPeg may be used for testing and learning how to
use the system. Place the Osstell TestPeg on the table or hold it in
your hand. Activate the Osstell Beacon by moving it and hold the
instrument tip approximately 3-5mm away from the top of the
Osstell TestPeg. The Osstell Beacon should start to measure and
present an ISQ value of 55 +/- 2 ISQ.
25087-00 ML
∅