LOW BATTERY INDICATION
The display will show
Lo
and need to be replaced.
AUTO SHUT-OFF
To extend battery life, the scale will automatically turn off
after approximately four minutes if no active weighing is
occurring. This feature is active with battery operation only.
CALIBRATION
For best results, calibrate the scale at regular intervals,
especially if using for prolonged periods. Calibration
weights are not provided with the scale.
CALIBRATION VALUES
MODEL #
CALIBRATION POINT (g.)
H-8109
200
H-8110
1,000
H-8111
2,000
The message
will appear
E
if the calibration steps are
not followed or the wrong
weight was used.
1.
Press and hold the
button until
7ENU
displayed. Release the
button, first menu
CAL
will be displayed.
NOTE: The calibration process can be aborted
by pressing
g
2. Press
to start span calibration.
0
3. The display shows
0
With no weight on the
pan, press the
0
button to establish the
zero point. The display
shows
while the
--C--
scale stores the zero
load signal.
4. The display shows the span calibration point, for
example, 200g. Place the specified calibration
weight on the pan and press the
display shows
--C--
calibration point signal.
PAGE 3 OF 9
OPERATION CONTINUED
when the batteries are weak
g
is
.
button. The
0
while the scale stores the
5. The display shows
[ -dONE-]
to the normal weighing
mode.
BACKLIGHT
The backlight settings can
be adjusted using the
menu. Press and hold the
button until
g
is displayed. Release the
button, first menu
be displayed. Press
select desired option.
MENU OPTION
b.Auto
b.ON
b.OFF
Once desired setting is displayed, press
and return to current weighing mode.
1-800-295-5510
uline.com
and returns
7ENU
will
CAL
to
g
Scale turns on light auto-
matically with any
button or weighing opera-
tion.
Backlight is always on.
Backlight is always off.
SETTING
to select
0
0621 IH-8109