ELECTROMAGNETIC COMPATIBILITY INFORMATION

Table 1
For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacture's declaration - electromagnetic emissions
The [KD-595] is intended for use in the electromagnetic environment specified below.
The customer or the user of the [KD-595] should assure that it is used in such an environment.
Emissions test
Compliance
RF emissions
Group 1
CISPR 11
RF emissions
Class B
CISPR 11
Harmonic emis-
sions
Not applicable
IEC 61000-3-2
Voltage fluctua-
tions/
Not applicable
flicker emissions
IEC 61000-3-3
Table 2
For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacturer's declaration - electromagnetic immunity
The [KD-595] is intended for use in the electromagnetic environment specified below. The customer or the user of the [KD-595]
should assure that it is used in such an environment.
IMMUNITY
IEC 60601test level
Compliance level
test
Electrostatic
discharge
± 6 kV contact
± 6 kV contact
(ESD)
± 8 kV air
± 8 kV air
IEC 61000-
4-2
Power
frequency
(50/60 Hz)
magnetic
3 A/m
3 A/m
field
IEC 61000-
4-8
112
Visiomed - KD-595 - Manuel d'utilisation - 072017
Electromagnetic environment - guidance
The [KD-595] uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely
to cause any interference in nearby electronic equipment.
The [KD-595] is suitable for use in all establishments other
than domestic and those directly connected to the public
low-voltage power supply network that supplies build-
ings used for domestic purposes.
Electromagnetic environment - guidance
Floors should be wood, concrete or ceramic tile. If floors
are covered with synthetic material, the relative humid-
ity should be at least 30 %.
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical commercial or
hospital environment.
Table 3
For all ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacturer's declaration - electromagnetic immunity
The [KD-595] is intended for use in the electromagnetic environment specified below. The customer or the user of the [KD-595]
should assure that it is used in such an environment.
IEC
IMMUNITY
60601test
Compliance level
Electromagnetic environment - guidance
test
level
RADIATED
3V/m 80
3 V/m
Portable and mobile RF communications equipment should be used no
closer to any part of the [KD-595], including cables, than the recommended
RF IEC
MHz to 2.5
separation distance calculated from the equation applicable to the fre-
61000-4-3
GHz
quency of the transmitter.
Recommended separation distance:
d = 1.2 √ P
80 MHz to 800 MHz
d = 2.3 √ P
800 MHz to 2,5 GHz
Where P is the maximum output power rating of the transmitter in watts
(W) according to the transmitter manufacturer and d is the recommended
separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey,a should be less than the compliance level in
each frequency range.b
Interference may occur in the vicinity of equipment
marked with the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios,
amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the
electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured
field strength in the location in which the [KD-595] is used exceeds the applicable RF compliance level above, the [KD-595]
should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary,
such as re-orienting or relocating the [KD-595].
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3V/m.
Visiomed - KD-595 - Manuel d'utilisation - 072017