3. Installation
3.1 Pump & Mattress Installation
NOTE- Unpack the box to inspect for any damage which may have occurred during shipment. If there
are any damages, please contact your dealer immediately.
1. Place the mattress or pad on top of the bed frame. Please note for the foot end. There are securing
straps on the base of the mattress. Secure the mattress firmly by fixing the straps to the bed frame,
ensure that moving sections of the bed frame are still free to move.
WARNING- The Pro-care 2 mattress must be applied on the underlying mattress.
2. Hang the pump onto bed rail (foot-end), and adjust hangers to best upright position of the pump, or
place the pump on a flat surface.
3. Connect air hose connectors from air mattress to the pump unit. When a "click" sound is felt or heard,
the connection is completed and secured.
NOTE- Check and ensure the air hoses are not kinked or tucked under mattress.
4. Plug the power cord into electrical outlet.
NOTE- 1. Make sure the pump unit is suitable for the local power voltage.
2. The plug is also served to disconnect the device.
5. Then turn the main power switch to ON position.
CAUTION- The pump can only be applied to the mattress recommended by the manufacturer.
Do not use it for any other purpose. (Applied part: air mattress) Do not position the
equipment so that it is difficult to operate the disconnecting device.
6. The cable management loops are provided along the sides of the mattress to cover the mains power
cable. Please make sure the wire has no signs of damage. After installation, the extra length of the
power cord, if any, should be neatly managed to avoid causing a trip and should be clear of moving bed
mechanisms or other possible entrapment area.
The EQUIPMENT should be firmly placed at position where users/doctors can access easily.