6�3 Blasting Nozzle Replacement; 6�4 Cleaning The Dosing Nozzle; 6�5 Lamp Cover; 6�6 Spare Parts - Renfert Basic master Traducción Del Manual De Instrucciones Original

Ocultar thumbs Ver también para Basic master:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 52
6.2
Condensation
• Inspect the water separator (14)�
• Drain any water off through the valve (15)�
6.3
Blasting Nozzle Replacement
• For installation see accessory supplement�
6.4
Cleaning the dosing nozzle
The dosing nozzle from the mixing chamber can become blocked
due to contamination and moist blasting sand�
For cleaning, the dosing nozzle is removed using the supplied tool
and replaced after cleaning�
• Vacuum clean the blasting tank completely�
• Hold the dosing nozzle with the tool and pull upwards (Fig� 11)�
• Clean the dosing nozzle (e�g� blow with compressed air)�
• Replace dosing nozzle�
Before remounting, the dosing nozzle should be completely dry!
• The dosing nozzle must be adjusted with blasting tanks of
25 - 70 µm (Fig� 12)!
see also supplement „Cleaning the dosing nozzle"
6.5
Lamp cover
In the course of time, the lamp cover can become matt� This cover can be ordered as a spare part�
6.6
Spare Parts
You can find components subject to wear and the spare parts on the spare part list in the internet at
www.renfert.com/p918.
The components excluded from the warranty (such as consumables or parts subject to wear and tear) are
marked on the spare part list�
Serial number and date of manufacturing are shown on the type plate of the unit�
6.7
Shipping Notes
• Empty all blasting tanks�
• Order Renfert transport packaging (Tel� +49 7731 8208-383)�
The return form can be downloaded from the „Advice" section on the Renfert website at
www.renfert.com�
- 9 -
14
15
Fig. 10
Fig. 11
Fig. 12
Abb� 12
EN

Publicidad

Tabla de contenido
loading

Tabla de contenido