EMC INFORMATION
The TP702 is suitable for use in the electromagnetic environment specified below.
The customer or the user of the TP702 should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to
80 MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to 2,5 GHz
NOTE 1
At 80 MHz and 800 Mhz, the higher frequency range applies.
NOTE 2
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects
and people.
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and
a
FM radiobroadcast, and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the TP702 is used exceeds
the applicable RF compliance level above, the TP702 should be observed to verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as re-orienting or relocating the TP702.
Over the frequency range 150 kHz to 80 MHz, field strengths are less than 3 V/m
b
Guidance and manufacturer's declaration – electromagnetic emissions
The TP702 is intended for use in the electromagnetic environment specified below.
The customer or the user of the TP702 should assure that it is used in such an environment
Emissions Test
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/Flicker
emissions
IEC 61000-3-3
20
TP702 EMC Information
Compliance
level
Portable and mobile RF communications equipment should be used no closer to
any part of the
3 Vrms
calculated from the equation appropriate for the frequency of the transmitter.
d = 1,2
3 V/m
d = 1,2
d = 2,3
where P is the maximum output power rating of the transmitter in
watts (W) according to the transmitter manufacturer and d is the recommended
separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site
survey,
Should be less than the compliance level in each frequency range.
Interference may occur in the vicinity of equipment marked with the following symbol:
Compliance
Electromagnetic Environment
Group 1
The TP702 uses RF energy only for its internal function. Therefore, its RF emissions are
very low and are not likely to cause any interference in nearby electronic equipment.
Class B
The TP702 is suitable for use in all establishments, including domestic establishments
and those directly connected to the public low voltage power supply network that supplies
buildings used for domestic purposes.
Class A
Complies
Electromagnetic environment - guidance
TP702
including cables, than the recommended separation distance
Recommended Separation Distance
P
P
80 MHz to 800 MHz
P
800 MHz to 2,5 GHz
a
OPERATOR'S MANUAL
TP702 T/PUMP
b
www.gaymar.com