RESISTANCES
If the «Voltest» light comes on, unplug the leads immediately. Present voltage
is upper to 15 V
To get the resistance in Ω, multiply the reading on the Ω scale (green) by coefficient
of the selected range: x1 or x100. The zero adjusting is automatic. To keep the
battery autonomy, avoid to untouch the rotary switch on the ranges.
Ω
Measurement extent
Internal resistance
End of scale current
Open circuit voltage
Accuracy
(3)
Permitted overload
1) Common to the continuity sound test
(2) In % at mid-scale
CONTINUITY SOUND TEST
Connection and specifications: idem Resistances
Place the selector switch on the
Continuous audible beep emitted for a resistance R < 80 Ω
BATTERY CHECKING
Put the rotary switch on the function
Short-circuit the leads, the needle must be positioned on the 0 (green square) of
Ω green scale. In the contrary, change the battery.
Additional uncertainty at the § 2 and § 3 uncertainty given.
TEMPERATURE
1 %/10°C for all the ranges except 2 %/10°C for 0.1 V DC and 2.5 %/10°C for AC
current ranges.
10 %/10°C in Ω.
FREQUENCY
Voltage ranges
- In the bandwidth 45 Hz to 500 Hz : 0 %, except 1000 V range: 0.5 %
- Outside the bandwidth 45 to 500 Hz and in the limits of § 3 bandwidth: 1.5 %,
except 10 V range: 4 %
Current ranges
- In the limit of § 3 bandwidth: 1.5 %
or
.
(1)
x1
5 Ω to 10 kΩ
140 Ω
10 mA
4. INFLUENCES
x100
500 Ω to 1 MΩ
14 kΩ
150 µA
-1.5 V
10 %
420 V
.
x1 function
x1 green.
15