Satelec PIEZOTOME Manual De Uso página 16

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 44
The practitioner has to choose the mode which is
most appropriate to the procedure.
The mode can be selected either:
• From the keyboard (Figure 8, #7): The user
should press the key corresponding to the mode
to be selected.
• Using the left footswitch (Figure 6, #1): Every
time the left footswitch is stepped on, the
device goes on to the next mode (1, 2, 3, 4,
1...)
An indicator light on the control screen confirms
the selection (Figure 7, #4).
The fluid flow rate associated with the selected
mode is shown on the LCD display (Figure 7, #1).
NB: the keys on the keyboard and the footswitch
button will be inactive if the device is emitting
ultrasound.
b) Irrigation settings
In P
function, the control screen LCD
IEZOTOME
display (Figure 7, #1) still shows the fluid flow
rate in ml/mn (on a white background). An
indicator light (Figure 7, #3) confirms the rate.
The user can adjust the fluid flow rate using the
+/- keys of the irrigation block on the keyboard
(Figure 8, #8).
The rate can be varied from 10-80 ml/mn in steps
of 5 ml/mn.
Irrigation can be activated either:
• Using the keyboard switch-key (Figure 8, #9)
• Using the footswitch switch-button (Figure 6,
#3)
If irrigation is On, the indicator (Figure 7, #2)
lights up.
If irrigation is Off, the indicator (Figure 7, #2)
goes out.
NB:
• the pump only operates while ultrasound is
being emitted.
• the keys on the keyboard and the footswitch
button are inactivated as long as the device is
emitting ultrasound.
c) Turning the ultrasound on
Ultrasound is turned On or Off using the
footswitch (Figure 6, #2). Every time the
footswitch is stepped on, the generator is
switched (On/Off).
• If irrigation is On, the indicator (Figure 7, #2)
lights up and the pump switches on at the same
time as the generator.
• If irrigation is Off, the indicator (Figure 7, #2)
goes out and the pump does not switch on at the
same time as the generator.
5. 5 NEWTRON FUNCTION
a) Power range
Range selection
In N
function, the N
EWTRON
keyboard is active (Figure 8, #5 and #6).
The device has 4 ultrasound power ranges. The
practitioner has to choose the range which is most
appropriate to the procedure.
The range can be selected:
• From the keyboard (Figure 8, #5): The user
should press the key corresponding to the range
to be selected.
• Using the footswitch button (Figure 6, #1):
Every time the button is stepped on, the device
goes on to the next range (G, Y, B, O, G, ...).
Range
Ultrasound Power
Green
Yellow
Medium
Blue
Orange
Very High
16
block on the
EWTRON
Procedure
Low
Periodontics mainly
Endodontics mainly
High
Prophylaxis mainly
Specific treatment
modalities

Publicidad

Tabla de contenido

Solución de problemas

loading

Tabla de contenido