Cleaning And Disinfection - HEINE mini 3000 Manual Del Usuario

Ocultar thumbs Ver también para mini 3000:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 21
Setting up the HEINE mini 3000
To setup the HEINE mini 3000
fresh batteries, unscrew the end cap (5) of the handle and insert new batteries (2 alkaline cells, LR6/AA) with the
plus contact facing the handle head.
To switch on the light source please slide the switch (4) down into position "1" . To switch off the light source slide
the switch (4) up to position "0" or clip it firmly over your coat pocket. The handle is then automatically switched off.
Performance Indicator: When turning on the mini 3000
to 100% and will then adjust to the brightness corresponding to the actual charge condition of the battery.
If you realize a significant drop of brightness of the LED immediately after switching on the instrument please
change or recharge the batteries
HEINE mini 3000
Otoscope operation
®
The otoscope may only be inserted into the auditory canal when a disposable (grey, 10) or re-usable (black, 9) tip is
fitted. Push the tip onto the instrument so that the internal projection engages in the slit (1). Twist to the right to lock.
The magnifying lens (2) magnifies 3x and can be swiveled aside for instrumentation. With the HEINE mini 3000
F.O. Otoscopes an insufflation bulb (7) can be fitted to the insufflation port (6). Therefore put the Insufflation port
connector (8) between insufflation port (6) and insufflation bulb (7).
The otoscope can be used with or without a tip for general illumination of e.g. the oral cavity or skin. Do not
contact sensitive or damaged skin.
The otoscope must only be used to examine the ear when a tip is fitted!
The contact time with skin has to be limited to max. 1 minute!
Pressure should be applied with great care when doing pneumatic testing of the tympanic membrane using
the insufflation bulb.

Cleaning and disinfection

Instrument head:
Clean the outside with a soft cloth, inside with a cotton wool bud, if necessary soaked in alcohol based disin-
fectant. Do not soak. Disinfection by spraying or immersion as well as steam sterilisation is not allowed. Please
only use disinfectants based on alcohol like Incidin
Disinfection by spraying or immersion as well as sterilization is not allowed and will damage the instrument!
Re-usable SANALON S tips:
These can be cleaned, disinfected and sterilized or boiled by any conventional method. Autoclaving is permitted.
AllSpec-Tips:
AllSpec-Tips are for single use only.
Do not attempt to clean them as the tip may be damaged and could injure the patient.
Battery handle:
The handle can be cleaned with a damp cloth (e.g. alkaline or pH- neutral detergent). For cleaning disinfection
of the outer surface we recommend the so called wipe disinfection by using detergents, which are released for
Medical Devices made out of plastic. Disinfection by spraying or immersion as well as sterilisation is not allowed.
Please use only disinfectants based on alcohol like Incidin
Disinfection by spraying or immersion as well as sterilization is not allowed and will damage the instrument!
Maintenance
HEINE mini 3000
LED Otoscopes do not require regular maintenance.
®
Changing the bulb:
With the HEINE mini 3000
With the HEINE mini 3000
wards. Push in the new bulb.
With the HEINE mini 3000
its clamp. Put in a new bulb and push it till its snaps.
10
Otoscopes
®
Otoscope, please screw the battery handle into the instrument thread. To enter
®
LED F.O. Otoscope the LED cannot be changed.
®
F.O. Otoscope, first unscrew the head from the handle and pull the bulb (11) down-
®
Otoscope, first rotate the magnifying lens (2) aside and pull-out the bulb (11) from
®
LED F.O. Otoscope the brightness will increase softly
®
Liquid and Bacillol
plus.
®
®
Liquid and Bacillol
®
plus.
®
®

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Mini 3000 f.o.Mini 3000 led f.o.

Tabla de contenido