Setting up the 2100 Bioanalyzer Instrument

To set-up the 2100 Bioanalyzer instrument:
1 Locate bench space that meets the requirements described in
2 Place the instrument on the bench in an upright position.
Damage by electrostatic discharge
C AU T I O N
The heater plate and electrodes are sensitive to electrostatic discharge.
3 Open the lid of the 2100 Bioanalyzer instrument and slide the electrode
space"
on page 67.
Make sure that you are grounded before inserting a chip to prevent the instrument
from being damaged by electrostatic discharge. The lid is grounded to leak off
electrostatic energy of the operator.
cartridge into the lid and move the lever into a flat (closed) position.
Installing the 2100 Bioanalyzer System
Setting up the 2100 Bioanalyzer Instrument