Voltage Tester Limit 110 / 112
General Specifications
- 110 indicates voltage 12, 24, 120, 230, 400,
690 VDC and VAC.
- 112 measure 12...690 V VDC and VAC.
Reading 1 V. Accuracy + 3 % + 5 digits.
- Continuity test 0...400 KΩ.
- Singel pole phase test 100...690 VAC.
- Rotary field indication 100...690 VAC. 50...60
Hz.
- Frequency 0...400 Hz.
- Automatisc detection.
- Acoustic signal.
- Power consumption Max 30 mA.
- Battery 2 pcs 1,5 V type AAA, LR03
- Temperatur range -10 – 55ºC.
- Humidity ranget 20%...96 %.
- Safety/Compliances CAT IV 600 V. IP 65.
Dubbel isolated. LVD (EN 61010-031:2002 ;
EN 61010-1:2002) EMC (EN
61326:1997+A1:1998+A2:2001+A32003)
Safety warning
To avoid possible electric shock or personal
injury, and to avoid possible damage to the
tester or to the equipment under test, adhere to
the following rules:
• Before using the tester inspect the function.
Do not use if it is damaged or part of the case
is removed.
• Do not apply more than the rated voltage.
Max 690 VAC.
• Keep alwavs your fingers behind the
measuring probes safety flange.
Voltage Test
1. Connect test probes L1 and L2 to the object
being measured.
2. Read voltage by LED lights (Limit 110) or
display shows voltage value (Limit 112).
DC voltage.
positive polarity, LED + illuminate, pos 11 (fig 1
LIMIT 110) alternative display shows – (Limit
112). When exchanced L1 and L2, LED –
illuminate (Limit 110) alternativt displayen
shows – (Limit112).
AC voltage.
(Limit110) alternativt display shows VAC (Limit
112).
If Probe L2 is connected to
Both LED + and – illuminate