Guidance And Manufacturers Declaration - Medtronic MMT-7306 Guia Del Usuario

Dispositivo usb
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 41
MP6025822-22C1 / A

Guidance and manufacturers declaration

Table 1. Guidance and Manufacturers Declaration - Electromagnetic Emissions
This CareLink USB (MMT-7306) is intended for use in the electromagnetic environment specified below. The customer or the
Emissions Test
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Table 2. Guidance and Manufacturers Declaration - Electromagnetic Immunity
This CareLink USB (MMT-7306) is intended for use in the electromagnetic environment specified below. The customer or the
IEC 60601 Test
Immunity Test
Electrostatic
±4 kV contact
discharge (ESD)
IEC 61000-4-2
Table 2, continued next page
8
user of the CareLink USB should make sure that it is used in such an environment.
Compliance
Group 1
Class B
Not applicable
Not applicable
user of the CareLink USB should make sure that it is used in such an environment.
Compliance
Level
Level
±4 kV contact The CareLink USB should not be affected by electrostatic discharge that
±8 kV air
±8 kV air
Electromagnetic Environment - Guidance
The CareLink USB uses RF energy only for its internal function. Therefore,
its RF emissions are very low and are not likely to cause any interference in
nearby electronic equipment.
The device is suitable for use in all establishments, including domestic
establishments and those directly connected to the public low-voltage
network that supplies buildings used for domestic purposes.
Electromagnetic Environment - Guidance
might occur under normal conditions of use.

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Carelink usb

Tabla de contenido