Wireless coexistence summary
The CareLink USB (MMT-7306) was tested against the data listed in the two tables for wireless
phones and wireless networks. The test results are described after the two tables.
Wireless phones
Telephones
WCDMA
TDMA
1700-2200 MHz
GSM
DECT

Wireless Networks

Networks
802.11b/g/n
Bluetooth
ZigBee
Data integrity: Each interference event was evaluated for current and latent anomalies. The
upload for each pump showed complete and uninterrupted functionality with no corrupted or
missing data.
Performance: The wireless functions operated safely and effectively in an intended use
environment.
Quality of service:
• In the absence of RF interference, all wireless transmissions were successfully executed.
6
Recommended Separation Distance
Frequency
800 - 960 MHz,
Recommended Separation Distance
Frequency
2.4 GHz
2.4 GHz
2.4 GHz
(meters)
0.3
0.3
0.3
0.254
(meters)
1
0.1
0.1
Recommended Separation Distance
(inches)
12
12
12
10
Recommended Separation Distance
(inches)
39
3.93
3.93