Table 2. Guidance and Manufacturers Declaration - Electromagnetic Immunity
This CareLink USB (MMT-7306) is intended for use in the electromagnetic environment specified below. The customer or the
IEC 60601 Test
Immunity Test
Electrical fast
±2 kV for power
transient/burst
supply lines
±1 kV for input/
IEC 61000-4-4
output lines
Surge
±1 kV line(s) to
IEC 61000-4-5
±2 kV line(s) to
Voltage
dips, short
0% 0.5 cycles
interruptions
and voltage
variations on
70% 25 cycles
power supply
0% 250 cycles
lines
IEC 61000-4-11
Power
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
Table 2, continued next page
user of the CareLink USB should make sure that it is used in such an environment.
Compliance
Level
Level
Not applicable
±1 kV for
input/output
lines
±1 kV line(s) to
line(s)
line(s)
±2 kV line(s) to
earth
earth
0% 0.5 cycles
0% 1 cycle
0% 1 cycle
70% 25 cycles
0% 250 cycles
3 A/m
3 A/m
Electromagnetic Environment - Guidance
For use in a typical domestic, commercial, or hospital environment.
Power frequency magnetic fields should be at levels characteristic of a
typical location in a typical commercial or hospital environment.