Malfunction
Li-ion battery cannot be loaded
9.

Technical Service

Risk of injury and/or malfunction!
Do not modify the product.
WARNING
Modifications carried out on this technical equipment may result in loss
of guarantee/warranty rights and forfeiture of applicable licenses.
Service addresses
Albert Kerbl GmbH
Felizenzell 9
84428 Buchbach / Germany
Phone:
+49 (0) 8086 933-100
Fax:
+49 (0) 8086 933-500
Email:
info@kerbl.com
www.kerbl.com
Other service addresses can be obtained from the address indicated above.
Detection
Li-ion battery plugged in, charge
state indicator does not light up
Cause
Charger contacts dirty
Charger contacts damaged
Li-ion battery defect
Charger defect
Too high a temperature is mea-
sured in the Li-ion battery during
the charging process
Battery charging error or Li-ion
battery defect
Remedy
Clean the contacts in the charging
bay, see Reprocessing
Send charger to manufacturer for
repair, see Technical Service
Send Li-ion battery to manufac-
turer for repair, see Technical Ser-
vice
Send charger to manufacturer for
repair, see Technical Service
Remove the Li-ion battery from
the charging bay, leave to cool and
repeat the charging process
If the problem occurs again, send
li-ion battery to manufacturer for
repair, see Technical Service
Remove the Li-ion battery from
the charging bay and repeat the
charging process
If the problem occurs again, send
li-ion battery to manufacturer for
repair, see Technical Service