Safety : Meets EN61010-2-032, UL61010B-2-032, IEC61010-1 2nd Ed., EN61010-1 2nd
Ed., UL61010-1 2nd Ed. CAT III-600 Volts AC & DC; Pollution degree : 2
EMC: Conforms to EN61326-1.
This product complies with requirements of the following European Community
Directives: 89/ 336/ EEC (Electromagnetic Compatibility) and 73/ 23/ EEC (Low Voltage)
as amended by 93/ 68/ EEC (CE Marking). However, electrical noise or intense
electromagnetic fields in the vicinity of the equipment may disturb the measurement
circuit. Measuring instruments will also respond to unwanted signals that may be
present within the measurement circuit. Users should exercise care and take appropriate
precautions to avoid misleading results when making measurements in the presence of
electronic interference.
ELECTRICAL SPECIFICATIONS
DC Voltage
Range
400.0 mV
4.000, 40.00, 400.0 V
600 V
NMRR:
CMRR:
Input Impedance:
Transient protection:
AC Voltage (50Hz ~ 500Hz)
Range
4.000, 40.00, 400.0 V
600 V
CMRR:
Maximum Crest Factor:
Input Impedance:
Transient protection:
ACD-14 Plus:
ACD-14 TRMS Plus:
Accuracy at 23 °C ± 5 °C & < 75% R.H.
Accuracy
±( 0.3% rdg + 4 digits)
±( 0.5% rdg + 3 digits)
±( 1.0% rdg + 4 digits)
>50 dB @ 50/60Hz
>120 dB @ DC, 50/60 Hz, Rs=1 kΩ
10 MΩ, 30 pF nominal (1000 MΩ for 400.0 mV range)
6.5 kV (1.2/50 μs surge)
Accuracy
±( 1.5% rdg + 5 digits)
±( 2.0% rdg + 5 digits)
>60dB @ DC to 60 Hz, Rs=1 kΩ
< 1.75 : 1 at full scale & < 3.5 : 1 at half scale limited to
fundamental and harmonics, that fall within the meter
specified AC bandwidth for non-sinusoidal waveforms
10 MΩ, 30 pF nominal
6.5 kV (1.2/50μs surge)
Average Sensing
True RMS sensing - 5% to100 % of range
11