ACA Current (Clamp-on 50Hz / 60Hz)
Range
40.00, 400.0, 600 A
Overload Protections:
ACD-14 Plus:
ACD-14 TRMS Plus:
1) Max Induced error from adjacent current carrying conductor: 0.05 A
2) Specified accuracy is from 1% rdg to 100% rdg of range and for measurements
made at the jaw center. When the conductor is not positioned at the jaw center,
position errors introduced are: Add 2% rdg to specified accuracy for measurements
made BEYOND jaw marking lines (toward jaw opening)
3) Add 8 digits to specified accuracy @ reading < 10% rdg of range
Frequency
Function
400.0 mVac
4.000 Vac
4.000, 40.00 Vac
400.0 Vac
600 Vac
Display counts: 5000
Resolution: 0.001Hz
Transient protection : VAC input jacks : 6.5kV (1.2/50μs surge)
Ohms
Range
400.0 Ω
4.000, 40.00, 400.0 kΩ
4.000 MΩ
40.00 MΩ
Open Circuit Voltage : 0.4 VDC typical
Transient protection :
Accuracy
1) 2) 3)
±( 1.5% rdg + 8 digits)
ACA Clamp-on jaws: 600 A rms continuous
Average Sensing
True RMS sensing - 10 % to 100 % of range
Sensitivity (Sine RMS)
350mV
1V
32V
90V
500V
Accuracy
±( 0.8% rdg + 8 digits)
±( 0.6% rdg + 4 digits)
±( 1.0% rdg + 4 digits)
±( 2.0% rdg + 4 digits)
6.5 kV (1.2/50μs surge)
Range
10 Hz ~ 2 kHz
5 Hz ~ 5 kHz
5 Hz ~ 100 kHz
5 Hz ~ 10 kHz
5 Hz ~ 5 kHz
12
Accuracy
±( 0.5% rdg + 4 digits)
±( 0.5% rdg + 4 digits)
±( 0.5% rdg + 4 digits)
±( 0.5% rdg + 4 digits)
±( 0.5% rdg + 4 digits)