Activating The Info Mode; Activating The Quick Menu; Menu Navigation - Dräger Polytron 8100 Instrucciones De Uso

Ocultar thumbs Ver también para Polytron 8100:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 128
4.3

Activating the Info mode

No password required.
The Info mode is used to show instrument relevant information.
This will not interrupt the normal operation of the instrument.
Tap and hold [UP] for more than 3 seconds.
The instrument information is shown on several screens.
No changes can be made.
Tap [UP] / [DOWN] to switch between the screens.
The Info mode can be exited at any time by tapping [OK].
If no key is tapped within 30 seconds, the instrument will
automatically return to normal operation.
Example of the Info mode
Depending on sensor the screens may vary:
1. Screen - Instrument information:
07.11.2003
12:34
SW Version : XX
Part No.
: 8317778
Serial No.
: ARUA0001
DeviceCode: 00006317
1/3
Instrument Info
A001.eps
Date and Time
Software version
Part number of the instrument electronics
Serial number of the instrument electronics
Instrument code
2. Screen - Sensor information:
Sensorname: O
2
Part No.
: 6809630
Serial No. : XXXXXXX
EEPROM Typ : 1
EEPROM Vers.: 1
2/3
Sensor Info
A002.eps
Installed sensor
Part number of the sensor
Serial number of the sensor
Software version
Software revision
Dräger Polytron 8100
3. Screen - Sensor configuration:
Gasname
: O
2
Range
: 25.00 Vo
4–20 SP
: 25.00 Vo
Alarm A1
: 19.00 Vo
Alarm A2
: 23.00 Vo
3/3
Sensor Config.
A003.eps
Selected gas
Maximum measuring range and units for the sensor
Measuring range and units for the analog interface. Only
shown if the 4 to 20 mA interface is enabled.
A1 Alarm set point and units
A2 Alarm set point and units
4.4

Activating the Quick Menu

No password required.
Tap and hold [DOWN] for 1 to 2 seconds.
The selected information about status or instrument
settings can be retrieved (e.g. warning messages, error
messages, installed modules and calibration data, time/
concentration graph, see Section 5.8.5 on Page 23.
No changes can be made.
4.5

Menu navigation

Scrolls upwards
Sets values
Scrolls downwards
Sets values
Confirms entry
Selects menus and functions
NOTICE
i
i
The instrument is designed for the magnetic wand
(part number 4544101: blue body, white logo) to be
used with the enclosure lid in place. If the enclosure lid
is not in place, the magnetic wand may activate two or
more buttons at once (cross-talk).
Operation
15

Publicidad

Tabla de contenido
loading

Tabla de contenido