Cable Junction Box PR 6130/38S
4.3
Cable gland
EN-12
Screen
yellow
The cable glands each have one sealing ring.
The cables have to be fed into the device via glands to ensure leak-tightness. The
following cable diameters are suitable: 8...13 mm for gland M20 and 3...6 mm for cable
gland M12.
The cable wires are connected to the terminals inside the device.
NOTICE
Property damage is possible.
If a seal insert is not used, it must be sealed with a supplied locking pin.
NOTICE
Property damage is possible.
Regularly check the itted cable gland for tightness and re-tighten it, if necessary.
Terminal contact
- supply voltage (connection cable)
IND
Terminal contact S - sense (connection cable – black)
Terminal contacts
Screen (connection cable)
1–2
Terminal contacts
Screen (load cells)
1–9
4 Installation and connection information