SHAKERS & ROCKERS
SSM1, SSM3, SSM4, SSM5, SSL1, SSL2,
SSL3, SSL4, SSL5
1. Introduction
Thank you for purchasing this piece of Stuart
equipment. To get the best performance from the
equipment please read these instructions carefully
before use.
Before discarding the packaging check that all
parts are present and correct.
For your own safety and that of others
please read and understand the safety advice
given below before using the equipment.
2. Safety Advice before use
If the equipment is not used in the manner
described in this manual and with
accessories other than those recommended by
Bibby Scientific the protection provided might be
impaired.
This equipment is designed to operate under the
following conditions: -
❖ For indoor use only
❖ Use in a well ventilated area
❖ Ambient temperature range +5°C to +40°C
❖ Altitude to 2000m
❖ Relative humidity not exceeding 80%
❖ Mains supply fluctuation not exceeding 10%
❖ Over-voltage category II IEC60364-4-443
❖ Pollution degree 2
❖ Use with a minimum distance all around of
200mm from walls or other items
The unit should be carried using both hands.
Do not use in a hazardous atmospheres or with
hazardous materials.
Ensure that the load is balanced.
Do not attempt to stop movement by hand.
Never move or carry the unit when in use or
connected to the mains electricity supply.
2
In the case of mains interruption, the unit will
not restart on restoration of the electricity supply.
In the case of mechanical interruption, (e.g.
motor stall), the unit will continue operating on
removal of interruption.
Mechanical energy can lead to breakage of glass
vessels. Use with care.
Mechanical hazard SSL2. Cradle to be
assembled onto platform before
operating.
3. General Description
Stuart shakers and rockers are ideal instruments
for mixing and aerating various biological and
chemical samples in the laboratory. All models
have digital setting of both speed and an
adjustable digital timer that automatically stops
after the set time (1 to 9999 minutes, 1 second
to 90 minutes or 1 minute to 9 hours). The SSM
range comprise of personal units with a small
footprint while the SSL is the larger labscale
equivalent.
Model SSM1 and SSL1 provides a smooth
uniform circular motion with an orbit of 16mm.
Speed range is 30 to 300rpm. Max load 3kg and
10kg respectively.
Model SSL2 has a side-to-side
reciprocating action ideal for vigorous
shaking applications. Speed range is 25 to
250rpm. Max load 10kg*.
IMPORTANT
* Maximum speed with 10kg load is 200rpm.
Maximum load for a speed setting of 250rpm is
5kg.
Model SSM3 and SSL3 provides a 3D gyratory
motion, ideal for low foaming agitation, DNA
extractions, staining and de-staining procedures
etc. Speed range is 5 to 70 rpm. Max load 3kg
and 10kg respectively. The angle of tilt can be
adjusted manually, without the need for tools,
between 3 and 12° (see section 4.6 for
instructions).
Model SSM4 and SSL4 have a see-saw rocking
action that creates a wave motion within vessels
such as culture flasks, Petri dishes etc. Speed