1.3

Display and controls

Displays
4
5
1
Electrode condition
Slope: more than 95%
and offset: ± (0-15) mV
Electrode condition is good
2
Measurement icon -
reading is locked, no matter measurement
3
Calibration icon -
4
pH/mV reading or slope
5
Error message - Err
6
Auto temperature compensation - ATC ; Manual temperature compensation - MTC
7
Temperature during measurement or Offset (mV) in calibration
1
2
6
Slope: 90-95%
or offset: ± (15-35) mV
Electrode condition is
acceptable
,
measurement or calibration is running; disappear means the
1
2
,
point or
point calibration in progress
(%)
in calibration
3
7
or offset: ± (35-60) mV
Electrode condition is not
good or needs cleaning
EN-3
Slope: less than 90%