STRETCHER CLEANING

These instructions are intended to provide recommended cleaning methods for the Stryker Model 1037 Trauma
Stretcher .
RECOMMENDED CLEANING METHOD
Note: Follow the cleaning solution manufacturer's dilution recommendations exactly .
• Remove the mattress prior to washing the unit; do not wash the mattress with the stretcher .
• Wipe the unit with cleaning solution and water per manufacturer's recommended dilution .
• Dry thoroughly . Do not replace the mattress on the stretcher until the unit is completely dry .
CAUTION
Before returning the unit to service after cleaning, ensure that the unit is functioning properly by verifying that all labels
are intact, raise/lower the stretcher height, brake/steer pedal locks properly in both positions, latch/unlatch the siderails,
raise/lower the fowler and gatch, and check all components for proper lubrication .
DO NOT STEAM CLEAN, PRESSURE WASH, HOSE OFF OR ULTRASONICALLY CLEAN THE STRETCHER .
Using these methods of cleaning are not recommended and may void this product's warranty .

RECOMMENDED CLEANERS

Suggested cleaners for stretcher surfaces:
Quaternary Cleaners (active ingredient - ammonium chloride) .
Phenolic Cleaners (active ingredient - o-phenylphenol) .
Chlorinated Bleach Solution (5 .25% - less than 1 part bleach to 100 parts water) .
Avoid over saturation and ensure the product does not stay wet longer than the chemical manufacturer's guidelines
for proper disinfecting .
CAUTION
Some cleaning products are corrosive in nature and may cause damage to the product if used improperly . If the products
suggested above are used to clean Stryker patient handling equipment, measures must be taken to ensure that the
stretcher is wiped with a damp cloth soaked in clean water and thoroughly dried following cleaning . Failure to properly
rinse and dry the stretcher will leave a corrosive residue on the surface of the stretcher, possibly causing premature
corrosion of critical components . Failure to follow the above directions when using these types of cleaners may void this
product's warranty .
www.stryker.com
Cleaning
1037-009-006 REV D
Return To Table of Contents