1.4.1 WARNINGS:
• Read this manual thoroughly, and be familiar with its contents prior to using this equipment.
• Test this equipment prior to a surgical procedure. This product was thoroughly tested at the fac-
tory before shipment.
• Disconnect the unit from the electrical outlet prior to inspecting system components.
• The ViroVac™ system is only intended and suitable for the applications that are mentioned in
the operating instructions.
• The smoke evacuator produces a strong vacuum. Adjust the airflow and the position of the
inlet end of the wand or tubing to prevent patient injury and to prevent suction of surgical
materials and surgical specimens.
• If the smoke evacuator is activated while the airflow is set to a high speed, it may produce a
sudden, strong suction action. Check the airflow setting before activating the smoke evacu-
ator to prevent patient injury and to prevent suction of surgical materials and surgical
specimens.
• To maximize patient safety, the tubing or wand should not come into direct contact with
tissue. Otherwise, patient injury may result.
• The BUFFALO FILTER ViroSafe® filters and single-use accessories are completely disposable.
Please dispose of according to your local codes or regulations and hospital policy. These filters
may be disposed of or incinerated, whichever is appropriate for your institution.
• Care should be taken to route the power cord, foot pedal, smoke evacuation tubing, and
EZ Link Remote Activation Device cable as to not cause a tripping hazard or crimping of cords.
• Do not operate this device in the presence of flammable or explosive gases.
• To avoid risk of Electric Shock, this equipment must only be connected to a supply mains with
protective earth.
• This equipment is intended for use by healthcare professionals only. This equipment may cause
radio interference or may disrupt the operation of nearby equipment. It may be necessary to
take mitigation measures, such as re-orienting or relocating the ViroVac™ or shielding the loca-
tion.
• The use of ACCESSORIES other than those specified by BUFFALO FILTER, or sold by BUF-
FALO FILTER as replacement parts for internal components, may result in increased emissions
or decreased immunity of the ViroVac™.
• This equipment should not be used adjacent to or stacked with other equipment and that if
adjacent or stacked use in necessary, the ViroVac™ should be observed to verify normal opera-
tion in the configuration in which it will be used.
Buffalo Filter ViroVac™ Operator's Manual: Page 6