Gima ET-100B Manual Del Usuario página 6

Termómetro digital de oído por infrarrojos
Ocultar thumbs Ver también para ET-100B:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

All manuals and user guides at all-guides.com
6
ENGLISH
The ear thermometer is a device capable
Probe lens
o f a c h i e v i n g i n f r a - r e d t e m p e r a t u r e
measurement when placed in the auditory
canal of a subject. It is a safe means of
measuring human body temperature through
Scan button
the ear, this product is for home and hospital
use, operator shall be at least 11 years old
and patient can be operator.
On/recall button
Specification
This appliance conforms to the standard of
ISO 80601-2-56:2017/AMD 1:2018 Medical
electrical equipment — Part 2-56: Particular
Battery cover
requirements for basic safety and essential
performance of clinical thermometers for
LCD
body temperature measurement. This
device complies with the requirements
of IEC 60601-1,IEC 60601-1-11 and IEC
60601-1-2.
Range of displayed temperature: 34.0°C ~ 44.0°C
Display L°C, when the temperature is under 34,0°C
Display H°C, when the temperature is over 44,0°C
Operating ambient:
Temperature: 16°C ~35°C
Humidity: <= 85% RH
Atmospheric pressure: 700~1060hPa
Altitude: <2000m
Storage and transport condition:
Temperature: from -25°C to 55°C
Humidity: <= 85% RH
Display resolution temperature range: 0.1°C
Accuracy: ±0.2°C (from 35.5°C to 42.0°C ) /
± 0.3°C (from 34.0°C to 35.5°C, from 42.0°C to 44.0°C)
Display: liquid crystal display, 3,5 digits
temperature unit: centigrade
display of memory: last ten memories
low voltage warning: the LCD display
Power consumption: 0.5 milliwatt in measurement mode
Battery: one 3V Lithium battery (CR 2032)
Battery life: 2500 takes or 4 months with 3-min. usage per day
Shelf life: 3 years
Dimension: 114mm x 35mm x 40mm
Net weight: 36g
Beeper sign: on/off, measuring finish, etc.
Self-testing sequence: Press the 'on/recall' button to turn on the thermometer and all of
symbol (See Fig. A) should be displayed on the LCD in one second.
Calibration frequency: If used privately no calibration is needed. For professional use it is
recommend to check once 2 year.
Manufacturing and Calibration date: See the label in the battery box.
Mode of operation of the clinical thermometer: adjusted mode
Reference body site: oral

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

25579

Tabla de contenido