Servicing - VWR VP 820 Manual De Instrucciones

Ocultar thumbs Ver también para VP 820:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

 Switch on pump with mains switch (see fig. 2)
Switching off the pump/removing from operation
 When transferring aggressive media, flush the pump prior to switch-off to increase the service life of the
diaphragm (see chapter 8.2.1).
 Switch off pump with mains switch (see fig. 2).
 Open pressure and suction lines to normal atmospheric pressure.
 Disconnect the power source.
8.

Servicing

8.1
Servicing Schedule
Component
Pump
Diaphragm and valve plates/sealings
Tab. 5
8.2
Cleaning
When cleaning, make sure that no liquids enter the inside of the housing.
8.2.1
Flushing Pump
 Before switching off the pump, flush it with air (if neccesary for safety reasons: with an inert gas) for
about five minutes under atmospheric conditions (ambient pressure).
8.2.2
Cleaning Pump
 Only clean pump with a damp cloth and non-flammable cleaning agents.
8.3
Changing Diaphragm and Valve Plates/Sealings
Conditions
Pump is switched off and mains plug is removed from the socket
Pump is clean and free of hazardous materials
Tubes removed from pump's pneumatic inlet and outlet
Spare part/tool
Spare part/tool
Service Set (according to chapter 10)
Philips-head screwdriver No. 2
Soft pencil
Tab. 6
Information on procedure
 Always replace diaphragm and valve plates/sealings together to maintain the pump performance.
With multi-head pumps, parts of the individual pump heads can be confused.
 Replace the diaphragm and valve plates/sealings of the individual pump heads consecutively.
Health hazard due to dangerous substances
in the pump!
Depending on the substance transferred, caustic burns or poisoning are possible.
 Wear protective clothing if necessary, e.g. protective gloves.
WARNING
 Flush pump before replacing the diaphragm and valve plates/sealings (see chapter 8.2.1).
Servicing interval
Regular inspection for external damage or leaks
Replace at the latest, when pump output decreases