The large Slim HP Color Cap is for use by recipients
older than 3 years of age, as this color cap and the Slim
HP magnet may become displaced upon dropping.
It is important to have the correct magnet strength, so
you do not experience discomfort or retention issues.
If magnet strength is insufficient, the headpiece may
fall off more than is acceptable. If magnet strength is
excessive, you may experience irritation or discomfort.
Consult your cochlear implant professional if there
are any concerns regarding magnet strength. If
deemed appropriate, an audiologist may change
the strength of the magnet in the headpiece. Do not
change magnet strength unless under the direction
of a cochlear implant professional. If you experience
any redness, irritation, or discomfort, discontinue use
of the headpiece immediately and contact a cochlear
implant professional. See the headpiece instructions
for use for additional information regarding adjustment
of headpiece magnet strength.
Store additional headpieces away from items with
magnetic strips (e.g., credit cards, hotel room keys,
etc.) as they may de-magnetize cards.
The digitally-coded, inductive transmission technology
used in this device is extremely reliable and experiences
virtually no interference from other devices. It should be
noted, however, that when operating the device near a
computer terminal or other strong electromagnetic field
(e.g., RFID system, metal detectors, electromagnetic
anti-theft systems), it may be necessary to be at least
24" (60 cm) away to ensure proper operation. If the
5