Cleaning And Storing The Device And Cuff; Error Messages / Troubleshooting; Specifications - Beurer BM 40 Manual De Instrucciones

Ocultar thumbs Ver también para BM 40:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 46
• To clear the memory of the relevant user memory, you
must first select a user memory. Start a request for ei-
ther average values or individual measured values and
press and hold the SET button for approx. 3 seconds.
All the values in the current user memory are deleted
after 3 brief beeps are output.
• You have the option to delete individual measured val-
ues by pressing the SET button when the measured
values are displayed after the measurement.
• To switch off, press the START/STOP button
• If you forget to switch off the device, it will switch off
automatically after 3 minutes.

8. Cleaning and storing the device and cuff

• Clean the device and cuff carefully using a slightly damp
cloth only.
• Do not use any cleaning agents or solvents.
• Under no circumstances hold the device and cuff under wa-
ter, as this can cause liquid to enter and damage the device
and cuff.
• If you store the device and cuff, do not place heavy objects
on the device and cuff. Remove the batteries. The cuff line
should not be bent sharply.

9. Error messages / troubleshooting

In case of faults, the
Error messages can occur when
1: Systolic pressure is not detected,
2: Diastolic pressure is not detected,
message appears in the display.
3: The cuff is fastened too tightly or loosely,
4: The pump pressure is higher than 300 mmHg,
5: There is a system or unit error.
In the above cases, you must repeat the measurement. Make sure
that the cuff tube is properly inserted and that you do not move or
talk. Re-insert the batteries if necessary, or else replace them.

10. Specifications

Model no.
.
Measurement
method
Measurement
range
Display accuracy Systolic ± 3 mmHg, diastolic ± 3 mmHg,
Measurement
inaccuracy
Memory
Dimensions
Weight
Cuff size
Permissible oper-
ating conditions
27
BM 40
Oscillometric, non-invasive blood pressure
measurement on the upper arm
Cuff pressure 0 – 300 mmHg,
systolic 60 – 280 mmHg,
diastolic 30 – 200 mmHg,
Pulse 30 –180 beats/minute
pulse ± 5 % of the value shown
Max. permissible standard deviation
according to clinical testing:
systolic 8 mmHg /diastolic 8 mmHg
2 x 60 memory spaces
W 119 mm x L 109 mm x H 60 mm
Approximately 387 g (without batteries,
with cuff)
22 to 35 cm
+ 10 °C to + 40 °C, 15-93 % relative air hu-
midity (non-condensing)

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

65815

Tabla de contenido