Dc Current; Ac Currents (Ac And Ac+Dc) - Metrix MX 556 Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Chapter II
Curve showing typical measurement error (5 V, 50 V, 500 V ranges)
Vin = 100 % F.S.
50000 pts
Vin = 50 % F.S.
25000 pts
Vin = 15 % F.S.
7500 pts
Vin = 5 % F.S.
2500 pts
1000 pts
1 kHz
6.3.

DC current

Selector
switch
position
µA mA
10 A
* See fuse specifications, § 7.1.1.
Number of points:
Range selection:
Additional error in Pk+/- mode for a pulse width of ≥ 1 ms: 1% R ± 50 D
For measurements performed on alternative signals, the selected range must tally with the
max. value of the signal peak.
6.4.

AC currents (AC and AC+DC)

Ranges
5 % to 100 % of range
500 µA
5 mA
50 mA
500 mA
10 A
1 % R + 30 D up to 2 kHz
* See fuse specifications, § 7.1.1.
Benchtop multimeter
Ranges
Accuracy
500 µA
0.2%R + 5 D
5 mA
0.2%R + 2 D
50 mA
0.05%R+2 D
500 mA
0.2%R + 2 D
10 A
0.5%R+ 5 D
Accuracy
40 Hz to 5 kHz
0.75 % R + 30 D
0.6 % R + 30 D
0.7 % R + 30 D
1%
10 kHz
Max
voltage
Protection
drop
700 mV
600 V
700 mV
600 V
700 mV
600 V
1.5 V
600 V
500 mV
600 V
50 000 (or 5 000 - see § 3.3.)
Automatic or manual for the 500 µA, 5 mA,
50 mA, 500 mA ranges
Protection
Fuses*
600 VRMS
F1 + F2
600 VRMS
F1 + F2
600 VRMS
F1 + F2
600 VRMS
F1 + F2
600 VRMS
F2
2%
5%
10%
Fuses*
Resolution
F1 + F2
RMS
F1 + F2
RMS
F1 + F2
RMS
F1 + F2
RMS
F2
RMS
Resolution
10 nA
1 mA
100 nA
10 mA
1 µA
100 mA
10 µA
1 A
1 mA
100 kHz
10 nA
100 nA
1 µA
10 µA
1 mA
Max. Peak
45

Publicidad

Tabla de contenido
loading

Tabla de contenido