DeVilbiss IntelliPAP DV53 STANDARD PLUS CPAP Serie Guía De Instrucciones página 19

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Quick View Menu Layout
The Quick View Menu shows a summary of information which is reported in the SmartCode data. Information in the Quick
View includes: Number of days used more than 4 hours (or can be set to 5 hours by clinician), AHI (Apnea/Hypopnea
Index), High Leak percent (percent of time at leak >95 L/m), and Daily Usage (average hours per day used). Each
parameter can be viewed in any of four time periods, 1 day, 7 days, 30 days and 90 days.
The following chart shows the Quick View information that is available on your CPAP display if viewing this information has
been enabled by your provider. The Quick View menu shows a summary of your SmartCode data for 1, 7, 30, and 90 day
periods. Each of the time periods contains info for "Number of Days > x hrs", Apnea Hypopnea Index (AHI), High Leak %
(>95 L/m) and Average Daily Hours of while breathing use.
Diagram of CPAP displays
Days ≥ 4H
1d
1/ 1 (100%)
AHI
1d
31.5+
High Leak
1d
31+%
Daily Usage
1d
9.3+ hours
Exit Menu
Press 
SmartCode Data and Adherence Score Information
Your Healthcare provider may contact you to retrieve SmartCode data and Adherence Score information. To display the
SmartCode data, press the LEFT arrow key once. Use the UP arrow or DOWN arrow keys to change the SmartCode
reporting period. Press the LEFT arrow key again to display the Adherence Score.
Definitions
Not all of the features in the table below are available in all models; some are optional and some depend on what your
physician and/or home care provider has selected. The table shows the items you may see, and describes the feature
associated with each item. Also, some features are only available in certain modes or with the DeVilbiss SmartLink
Therapy Management Module – these are noted in the 'Mode' column below. If you do not see some of these options, your
model or confi guration does not support them.
Please note that the values displayed are examples only.
Display Example
OFF
Quick View Menu
Press 
P: 5.0
Heat:Off
Set Pressure
10.0
cmH2O
A-DV53
Days ≥ 4H
7d
7/ 7 (100%)
AHI
7d
31.5+
High Leak
7d
31+%
Daily Usage
7d
9.3+ hours
Description
Blower OFF, the CPAP unit is plugged in to AC or DC power with the
blower OFF.
Pressing the down key enters the Quick View menu. NOTE: The
"Quick View" menu is only shown if enabled from the clinical menu.
This is the default screen when the device is turned on. The actual
Auto
mask pressure and heater setting are shown. The displayed pressure
may vary slightly while breathing. If the heater system is not
connected, the heat setting will be blank.
This is a prescription setting which controls the operating pressure.
Days ≥ 4H
30d
30/30 (100%)
AHI
30d
31.5+
High Leak
30d
31+%
Daily Usage
30d
9.3+ hours
UP () DOWN () buttons
Days ≥ 4H
90d
90/90 (100%)
AHI
90d
31.5+
High Leak
90d
31+%
Daily Usage
90d
9.3+ hours
Mode
All
All
All
CPAP Mode
EN - 19

Publicidad

Tabla de contenido

Solución de problemas

loading

Este manual también es adecuado para:

Intellipap dv53Intellipap dv5hh

Tabla de contenido