Select The Location; Initial Calibration - OHAUS PR Serie Guia Del Usuario

Ocultar thumbs Ver también para PR Serie:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 9
EN-2
2. INSTALLATION

2.1 Select the Location

Avoid heat sources, rapid temperature changes, air current or excessive vibrations. Allow sufficient
space.
2.2 Leveling
Be sure the balance is level before it is used or after its location
is changed.
The PR balance has a level bubble in a small round window
beside the display.
To level the balance, adjust the 2 leveling feet until the bubble
is centered in the circle.
Please refer to the right figure for leveling.
2.3 Connecting Power and Acclimatising the Balance
Connect the DC output connector to the power receptacle on
the rear of the balance. Then connect the AC adapter plug to a suitable electrical outlet.
Acclimatising
It is suggested that the balance should not be used until it has been connected to power and
acclimatised to the environment for a certain period of time. In the case of a balance with the precision
above 0.1 mg, the acclimatisation time should be 1.5 hours; in the case of balance with the precision of
0.01 mg, the acclimatisation time should be more than 4 hours.
2.4 Connecting the Interface
The PR balance has an RS232 port.
Use the RS-232 port to connect either to a computer or a printer with a standard (straight-through) serial
cable.
Interface connections on the rear of the balance
RS232
RS232: Used to connect to PC or Printer
Note: See the Printing section for Connecting, Configuring and Testing the Printer / Computer Interface.

2.5 Initial Calibration

For best results, the instrument's calibration should be checked prior to first use. If adjustment is needed,
refer to the Calibration section of the instruction manual.

Publicidad

Tabla de contenido
loading

Tabla de contenido